USB-1608G Series

 

High-Speed Multifunction USB Devices

KEY HIGHLIGHTS

The USB-1608G Series of high-speed multifunction USB data acquisition (DAQ) devices are ideal for low to medium channel count applications, and provide analog along with digital and counter/timer functions. Up to two analog outputs are available. OEM versions are available for embedded applications.

ANALOG Input
Channels Resolution Max Sample Rate
16 SE/8 DIFF 16-bit Up to 500 kS/s
Sampling Ranges Isolation
Multiplexed ±10 V, ±5 V, ±2 V, ±1 V
ANALOG Output
Channels Resolution Speed
Up to 2 32-bit Up to 500 kS/s
Digital I/O
Channels Counter/Timers Encoder
8 2/1
Software Power
OS Support Drivers Power
Windows®, Android™, and Linux® Universal Library™ SW Suite Bus powered

Part Number Description Qty Price
USB-1608G USB-based 16-channel, 250 kS/s data acquisition device with eight digital I/O, two 32-bit counter inputs, and one timer output
$399.00
USB-1608GX USB-based 16-channel, 500 kS/s continuous sampling data acquisition device with eight digital I/O, two 32-bit counter inputs, and one timer output
$599.00
USB-1608GX-2AO USB-based 16-channel, 500 kS/s continuous sampling data acquisition device with two analog outputs, eight digital I/O, two 32-bit counter inputs, and one timer output
$799.00
USB-1608G-OEM USB-based DAQ board with 16-channels, 250 kS/s sample rate, two 32-bit counter inputs, one timer output, and eight digital I/O
$399.00
USB-1608GX-OEM USB-based DAQ board with 16-channels, 500 kS/s sample rate, two 32-bit counter inputs, one timer output, and eight digital I/O
$599.00
USB-1608GX-2AO-OEM USB-based DAQ board with 16-channels, 500 kS/s sample rate, two analog outputs, two 32-bit counter inputs, one timer output, and eight digital I/O
$799.00
Software
DAQami™ v4.2 Data acquisition companion software for acquiring data and generating signals
$49.00
DASYLab® LITE Icon-Based Data Acquisition, Graphics, and Control Software
More Options...
$499.00
DASYLab® BASIC Icon-Based Data Acquisition, Graphics, Control, and Analysis Software. Includes standard analysis modules and unlimited module count. - Most Popular $1,299.00
DASYLab® FULL Icon-Based Data Acquisition, Graphics, Control, and Analysis Software. Includes all DASYLab Basic features plus standard modules, 200 layout windows, unlimited module count, and control sequencer. $1,799.00
DASYLab® PRO Icon-Based Data Acquisition, Graphics, Control, and Analysis Software. Includes all DASYLab Full features plus advanced signal analysis and control modules. $2,499.00
DASYLab® RUNTIME DASYLab Runtime allows you to run an existing worksheet application on an additional computer (with compatible hardware configurations) $449.00
TracerDAQ® Pro Out-of-the-box virtual instrument suite with strip chart, oscilloscope, function generator, and rate generator – professional version
$199.00

Product Reviews


4.8 / 5
Ease of Use
4.4 / 5
Value
4.6 / 5
Techincal Support
4.6 / 5

97% of new reviewers recommend this product


USB-1608G
April 17, 2017
This product's documentation and quality is excellent.

I am very happy with my purchase of the USB-1608G DAQ. It was my first time purchasing a DAQ, and the documentation was very helpful. It was very easy to use with LabVIEW, and the ULx VI software is very good. I will recommend this product and this company in the future.

  • Job Title: Research Student
  • Industry: Academics
  • Application: Impact force measurement
  • From: Austin, TX
  • Company Size: 1-100
  • Software Used: NI LabView

  • Ease of use
  • Value
  • Techincal Support
Yes, I recommend this product!

Was this review helpful? Yes (0) / No (0)

USB-1608G
February 20, 2017
Good

Overall nce

  • Job Title: Research associate
  • Industry: TCS
  • Application:
  • From: Beaverton or
  • Company Size: 1-100
  • Software Used: C++/C#/VB

  • Ease of use
  • Value
  • Techincal Support
Yes, I recommend this product!

Was this review helpful? Yes (0) / No (0)

USB-1608G
February 14, 2017
Good device

I initially had some difficulties with the 1608G but I believe these were due to grounding issues. Once set up properly it has worked very well and is stable over temperature compared to previous USB products. Programming (VB6) has followed previous device code so it was easy to upgrade.

  • Job Title: President / Chief Battery Scientist
  • Industry: Battery
  • Application: Monitoring
  • From: Winfield, IL
  • Company Size: 1-100
  • Software Used: C++/C#/VB

  • Ease of use
  • Value
  • Techincal Support
Yes, I recommend this product!

Was this review helpful? Yes (0) / No (0)

USB-1608G
April 28, 2016
Surprisingly good and very cost effective

I work for a test and measurement company but when I needed a digitizer quickly I looked online and quickly homed in on Measurement Computing's products. I expected it to arrive and be a hassle to setup and use effectively. The opposite was true.

  • Job Title: Director
  • Industry: Test and Measurement
  • Application: Data Logging
  • From: Boul
  • Company Size: 1-100
  • Software Used: DAQami

  • Ease of use
  • Value
  • Techincal Support
Yes, I recommend this product!

Was this review helpful? Yes (2) / No (0)

USB-1608GX-2AO
January 25, 2016
A reliable product

We've purchased quite a number of this item, and they always operate reliably. We've used Measurement Computing products for many years, and whenever technical phone support is needed, it's always excellent.

  • Job Title: Director
  • Industry: Medical
  • Application: Infrared thermal imaging
  • From: Tampa, Florida
  • Company Size: 1-100
  • Software Used: C++/C#/VB

  • Ease of use
  • Value
  • Techincal Support
Yes, I recommend this product!

Was this review helpful? Yes (0) / No (0)

USB-1608G
January 04, 2016
Great product for great price

This DAQ device is easy to plug-n-play and begin acquiring accurate data with your application. It is inexpensive and very useful. We built a little protective box for ours as the application does not very much.

  • Job Title: Postdoctoral Fellow
  • Industry: Biomedical Research
  • Application: Pneumatic Control and Biosensor Recording
  • From: Rochester, MN
  • Company Size: 5000*
  • Software Used: NI LabView

  • Ease of use
  • Value
  • Techincal Support
Yes, I recommend this product!

Was this review helpful? Yes (0) / No (0)

USB-1608G
September 16, 2015
Highly versatile

We had an application that needed 16 fast/high resolution 16 bit analog inputs along with some analog outputs and some unique timing requirements. What I found especially interesting are the 2 PWM outputs which are extremely versatile. The PWN output was programmable across the required frequency range and allows user programmable duty cycle. We used the second timer output as the A/D trigger signal (external). These counters even allow one to specify an exact number of pulses to generate or "free running".

  • Job Title: Sr. Controls and software engineer
  • Industry: Electronics manufacturing
  • Application: Product laboratory test fixture
  • From: Chicago, IL
  • Company Size: 5000*
  • Software Used: C++/C#/VB

  • Ease of use
  • Value
  • Techincal Support
Yes, I recommend this product!

Was this review helpful? Yes (0) / No (0)

USB-1608G
August 24, 2015
Parts order for the lab

it was a great experience to buy from Measurement Computing we order parts for our lab here in South Florida. Thanks, Jeff

  • Job Title: Laboratory Manager
  • Industry: Education "Florida International University"
  • Application:
  • From: South Florida
  • Company Size: 5000*
  • Software Used: C++/C#/VB

  • Ease of use
  • Value
  • Techincal Support
Yes, I recommend this product!

Was this review helpful? Yes (0) / No (0)

USB-1608G
June 22, 2015
Easy transition from PCI to USB

Your USB-1608G has made it easy for our company to progress to USB A/D transition. We are now actively supplying your USB A/D as part of the Ariel Performance Analysis System. Thank you for your continued support. Sincerely,

  • Job Title: CEO
  • Industry: Medical
  • Application: Biomechanic Human Performance Analysis
  • From: California and NJ
  • Company Size: 1-100
  • Software Used: C++/C#/VB

  • Ease of use
  • Value
  • Techincal Support
Yes, I recommend this product!

Was this review helpful? Yes (1) / No (0)

USB-1608G
February 16, 2015
I was an exception replacement for an older unit.

I purchased this items as a replacement for a USB-1608fs. The program that I had setup for use with the older unit needed very few changes to work with the USB-1608G and I was back to testing in a few minutes. The differential inputs is a very useful upgrade that will allow more flexibility in testing. The analog output will allow for better simulations while developing new test equipment.

  • Job Title: Test Technician
  • Industry: Automotive
  • Application: Tools
  • From: Kenosha, WI
  • Company Size: 101-500
  • Software Used: DASYLab

  • Ease of use
  • Value
  • Techincal Support
Yes, I recommend this product!

Was this review helpful? Yes (0) / No (0)




Dear Sir, I tried to generate installer for my labview program based on NI recommendation which can be found @ http://kb.mccdaq.com/KnowledgebaseArticle50609.aspx when I try to install in a PC which does not have both labview and Instacal, the program start well, but not find USB-1608G hardware. even it does not power up the DAQ. What can I do? Regards,

  • Job Title: Electrical Engineeer
  • Industry: Technology
  • Application: Electric Meter
5 months ago
AI AA


Answers

In that FAQ, the created application installer is asked to install our InstaCal driver set. But the end-user must still manually run InstaCal (the user must launch inscal32.exe) so that the USB-1608G hardware is assigned a board number (usually board # 0). If the user cannot locate the file named inscal32.exe, then the InstaCal driver set was not installed properly by your installer.

5 months ago
Measurement Computing

Was this answer helpful? Yes (0) / No (0)

When I open and close the labview project I get "Dependency loaded from new path & .NET assembly changed" warnings. I can not generat .exe and installer for the project.

  • Job Title: Electrical Engineer
  • Industry: Technology
  • Application: Electric Meter
5 months ago
AI AA


Answers

These warnings are because we saved/distribute our VIs in an older version of LabVIEW than you are developing in. Simply re-save the VIs, and then the next time you open your project, you will not get these warnings.

5 months ago
Measurement Computing

Was this answer helpful? Yes (0) / No (0)

How is the analog ground/analog inputs isolated from the digital ground (and USB shield)? Are they shorted or is there a true isolation between them?

  • Job Title: Design engineer
  • Industry:
  • Application:
5 months ago
eebike


Answers

The USB-1608G series provides no ground isolation. The analog and digital circuits each have their own ground planes. This helps with noise reduction in the circuits. However, they eventually come to a single point connection under the ADC. Measurement Computing does offer DAQ products with isolation. Please contact us at 1-800-234-4232 Ext-3 to discuss the options.

5 months ago
Measurement Computing

Was this answer helpful? Yes (0) / No (0)

Dear Sir, How can I use DAQ Assistant for USB-1608G. Where can I find it? Regards,

  • Job Title: Electrical Engineer
  • Industry: Technologies
  • Application: Electric Meter
5 months ago
AI AA


Answers

Unfortunately, DAQ Assistant will only work with National Instrument’s LabVIEW and NI DAQ products. For help with programming your LabVIEW application using our ULx drivers for LabVIEW, please refer to our KnowledgeBase article “Getting Started with LabVIEW and ULx” here: http://kb.mccdaq.com/Search.aspx?Keywords=Getting%20started%20with%20LabVIEW.

5 months ago
Measurement Computing

Was this answer helpful? Yes (1) / No (0)

Hi, I'd like to know how one can realize a burst measurement of n<16 channels with Labview and ULx. The ULx help doesn't even mention the burst mode which was proclaimed in the datasheet.

  • Job Title: research engineer
  • Industry:
  • Application:
9 months ago
drr


Answers

Burstmode is not a setting/choice that is supported by ULx. However, specific to the USB-1608G series, note that when external clock mode is used, it mimic burstmode (i.e. all enabled channels are sampled once per clock pulse). So, if you want to use the on-board (internal clock), then the workaround is to instead wire the timer output of the unit over to its external clock input, specify external clock for the AI, and then start the timer output (using ULx’s counter output VIs) at the desired sample rate.

9 months ago
Measurement Computing

Was this answer helpful? Yes (0) / No (0)

The Retrigger Mode distinguishes the USB-1608G from similar devices e.g. USB-1608FS-Plus. The datasheet of USB-1608G says the Trigger Input is digital but below it states that the Retrigger Mode is sensitive to analog trigger events. Does that mean I can connect an analog trigger source to the TRIG input and set the sampling time via the trigger count such, that each time the a rising edge is detected at the TRIG input, the measurement is done?

  • Job Title: Student
  • Industry:
  • Application: Triggered Sampling
10 months ago
joe546 Bavaria


Answers

The trigger on the USB-1608G series accepts only TTL level, digital input (0 to 5 volt logic). The portion of the data sheet that I believe you are referring to describing the “Retrigger” feature… “In retrigger mode, you can set up repetitive analog input trigger events” is referring to the digital trigger event starting your “Analog Input Scan” so as to capture an analog event, triggered by the TRIG input. Unfortunately, you cannot connect an analog trigger signal. However, the “Retrigger” feature will work the way you have described below in that the trigger will rearm itself and wait for the next DIGITAL trigger event to occur to start another acquisition. Excerpt from the ULHELP: Retriggering Use RETRIGMODE with AInScan() to set up repetitive trigger events. Use the ConfigItem option BIADTRIGCOUNT with cbSetConfig() to set the A/D trigger count. When using RETRIGMODE, set the values for the Count argument (cbAInScan()/AInScan()) and the BIADTRIGCOUNT argument (cbSetConfig()/SetAdRetrigCount()) to an integer multiple of the number of channels in the scan. That way, the entire buffer, or the portion of the buffer defined by BIADTRIGCOUNT, will contain updated data.

10 months ago
Measurement Computing

Was this answer helpful? Yes (1) / No (0)

  • Job Title: instructor
  • Industry: Computer Tech.
  • Application:
13 months ago
bduman


Answers

The USB-1608G is not directly compatible with 4-20 mA signals. It lacks the necessary signal conditioning. However, with the use of a low cost, precision 249 ohm resistor placed across the USB-1608G’s +-5 volt differential inputs you can easily create a current shunt which will produce a 1-5 volt drop across the shunt. With 1 volt equal to 4 mA and 5 volts equal to 20 mA.

13 months ago
Measurement Computing

Was this answer helpful? Yes (0) / No (0)

  • Job Title: Senior Sensor Architect
  • Industry:
  • Application:
14 months ago
lightyear66 Folsom, CA


Answers

The USB-1608G is a multiplexed DAQ device using a single A/D converter. If you need a simultaneous DAQ device, please take a moment to review the following products: USB-1608FS-Plus, USB-1608HS, USB-1602HS, or the USB-1604HS DAQ device.

14 months ago
Measurement Computing

Was this answer helpful? Yes (1) / No (0)

I have USB-1608GX-2AO DAQ device which I am using with Datawave (Dwavetech.com) Experimenter Workbench. I want to use the Digital I/O to output TTL- like pulses to an Arduino board . When I use the Experimenter Workbench object for digital output to speci

  • Job Title: Lab Instructor
  • Industry: Neuroscience Research/ Education
  • Application: Electrophysiology/Optogenetics
30 months ago
smilway Memorial University of Newfoundland


Answers

You can test the digital output function by running the InstaCal Analog Loopback test. Close your 3rd party program before launching InstaCal – then you choose which digital output bit for it to toggle, and you wire (loopback) it to an analog input channel. You should see a square wave on the screen. The amplitude of the digital output is not adjustable. If the digital output toggles properly in InstaCal, then the issue is with your 3rd party program.

30 months ago
MeasurementComputing

Was this answer helpful? Yes (0) / No (0)

  • Job Title:
  • Industry:
  • Application:
30 months ago
EEwannabe Calif


Answers

Hello - I believe you are referring to NI’s SignalExpress. MCC’s ULX driver is not compatible with SignalExpress. ULx is compatible with the traditional LabVIEW graphical programming environment.

30 months ago
MeasurementComputing

Was this answer helpful? Yes (0) / No (0)

I need to measure an analog signal between 5 and 6 volts that needs to be accurate to 0.001 volt. Can the 1608G do this?

  • Job Title: Electrical Engineer
  • Industry:
  • Application:
32 months ago
Rhan Massachusetts, USA


Answers

Hello - The USB-1608G will not meet your accuracy requirement of 0.001 volts. The stated accuracy for the +- 10 volt input range is 4075 uV (4.075 mV) and for the +- 5 volt input range is 2266 uV (2.266 mV). To attain a 0.001 volt accuracy you would have to use one of our 24 bit DAQ devices. Like the USB-2404-10. http://www.mccdaq.com/products/product-listing.aspx?oid=1596&r=24 but please do not confuse accuracy with resolution. If you need 0.001 volts of resolution then the USB-1608G will work just fine. Your resolution for the +- 10 volt range would be 0.000305 volts/LSB. I have added a link below to a Knowledge-Base whitepaper on accuracy and resolution. Please do not hesitate to contact MCC with any additional questions. http://kb.mccdaq.com/KnowledgebaseArticle50043.aspx?Keywords=Accuracy

32 months ago
MeasurementComputing

Was this answer helpful? Yes (0) / No (0)

  • Job Title: University renewable energy Innovation department
  • Industry: University renewable energy Innovation department
  • Application: Acquiring temperature data for electrolyzer project
45 months ago
Xamoso Spain


Answers

No. The thermocouple generates a tiny voltage, and even with a 24 bit DAQ there is not enough voltage range on the thermocouple to measure it directly with the USB-1608G. Alternatives are to make a thermocouple pre-amplifier to boost the voltage up to a level where it can be directly read by the DAQ. Also DAQs that do measure thermocouples have internal pre-amplifiers, like the USB-2408 Series DAQ. The DAQs that do measure thermocouples also have an internal reference thermocouple.

38 months ago
zmonkey

Was this answer helpful? Yes (0) / No (0)

You should get a USB-TC or USB-TEMP for that. They are designed for reading thermocouples, and give temperature directly. Unlike the USB-1608g they compensate for the temperature-dependent voltage drop due to the mismatch between the terminal on the DAQ and the thermocouple wire.

44 months ago
Noxsc

Was this answer helpful? Yes (1) / No (0)

The USB-1608G was not designed to read thermocouples directly. It lacks 3 things needed: very high gain/amplification, cold junction compensation circuitry, and low pass filtering. Instead choose a product designed with thermocouples in mind, such as the USB-TC or USB-2408.

45 months ago
MeasurementComputing

Was this answer helpful? Yes (0) / No (0)

I was about ordering for USB-1608GX (plus C37FF-2, CIO-MINI37, ISO-5B32-01, ISO-RACK16 etc ) but realized that this device cannot be daisy-chained unlike USB-1616FS. We are about developing a Windows Application to acquire data from over 50 analogue sens

  • Job Title: Equipment Support Engineer
  • Industry: Oil and Gas (Mud-Logging / LWD)
  • Application: Drilling Data Acquisition for Mud-Logging
31 months ago
Emenyke Warri Nigeria


Answers

Hello Emenyke, Yes, the AI-EXP32 will expand the channel count on the USB-2416 to 64 single ended or 32 differential. My recommendation for a 50 channel analog system with signal isolation would be the Daqlab/2005 with the DBK42 x 4, ISO-5B32-01 modules and a CA-37-4 cable to make the interconnections to the Daqlab/2005 unit.

31 months ago
MeasurementComputing

Was this answer helpful? Yes (0) / No (0)

I need some isolation / signal-conditioning for connected sensors. Which of your Isolation Modules or Signal conditioners can go with this product if I am connecting a 4-20ma output Sensor transducer.

  • Job Title: Equipment Support Engineer
  • Industry: Oil and Gas (Mud-Logging / LWD)
  • Application: Mud-Logging Data Acquisition
31 months ago
Emenyke Warri Nigeria


Answers

For 4-20 mA isolation, MCC recommends the ISO-5B32-01 module. You can review the specifications here: http://www.mccdaq.com/signal-conditioning/ISO-5B32-01.aspx There will be other accessories required for you to interface the ISO-5B32-01 to your USB-1608GX DAQ device. I recommend that you purchase, along with the ISO-5B32-01 modules(s), an ISO-RACK08 backplane (http://www.mccdaq.com/signal-conditioning/ISO-RACK08.aspx), a CIO-MINI37 and a C37FF-2 interface cable. This setup will allow you to make the necessary connections between the ISO-5B32-01 module and your USB-1608GX. Another option would be to purchase our USB-2404-UI which supports channel to channel and channel to ground isolation as well as (4) +- 25 mA input ranges. http://www.mccdaq.com/usb-data-acquisition/USB-2404-UI.aspx

31 months ago
MeasurementComputing

Was this answer helpful? Yes (0) / No (0)

Question about the USB-1608G and USB-1608GX. I would like to use one unit to collect data for say for 10-seconds using a digital trigger and the other unit to continuously scan some other inputs at the same time and display them live.

  • Job Title: Business Owner
  • Industry: Die Casting
  • Application: Monitoring
41 months ago
Lglendening St. Joseph, MI, USA


Answers

Hi Lglendening - Yes, It is possible to use 2 or more of our products with the Universal Library API in this manner.

41 months ago
MeasurementComputing

Was this answer helpful? Yes (0) / No (0)

Hi! I plan to use a DAQ device to gather analog signal response to a specific signal output. The input and out data points does not have to be strictly simultaneous. For example, while outputing a step function with a time duration of 50ms, can the input channel acquire data during and after the 50ms?\ Meanwhile, may I get a rough dimension of the device? Thank you!

  • Job Title: Student
  • Industry:
  • Application:
7 months ago
Big Head


Answers

Yes, the USB-1608GX-2AO with the appropriate software will support concurrent analog input and output operations. The dimensions for the USB-1608GX-2AO are as follows: 5.0 (L) x 3.53 (W) x 1.40 (H). For future reference dimensions for our products can be located in our User’s Manuals. Please feel free to contact us if you have any additional questions.

7 months ago
Measurement Computing

Was this answer helpful? Yes (0) / No (0)

I am looking for a daq system that allows the following: - analog output at 250 kHz (single channel). -synchronized sampling on two analog input channels at 250 kHz (the samples for each channel need to correspond to the same points in time). - Starting the sampling of the analog inputs synchronized with the analog output. This is, I want to start sampling the analog input channels at the same time I start the analog output. Is this possible with this device?

  • Job Title: Professor
  • Industry:
  • Application:
10 months ago
Jerry546 Cincinnati, Ohio


Answers

Hello - The USB-1608GX-2AO does not support simultaneous sampling of the analog inputs nor does it support synchronized sampling of the analog inputs and outputs. For this you will need to use our USB-1602HS-2AO or USB-1604HS-2AO DAQ devices. http://www.mccdaq.com/usb-data-acquisition/USB-1602-1604-Series.aspx Both devices support simultaneous analog inputs and synchronized sampling of the analog inputs and outputs.

10 months ago
Measurement Computing

Was this answer helpful? Yes (0) / No (0)

Raw input mode: each of them can be read separately and instantaneously. Logic analyzer mode (sampling the 4 inputs at a fixed frequency and filling a buffer) Raw outputs mode: each of them can be set separately and instantaneously . Signal generator mode ( synchronized streaming of 1 to 4 outputs at a fixed frequency, using a buffer)

  • Job Title: trainee
  • Industry:
  • Application:
16 months ago
divya bangalore(india)


Answers

What you are describing here cannot be accomplished with using the digital input/output bits on the USB-1608GX-2AO. The digital I/O bits on the USB-1608GX-2AO are updated via software and not hardware paced with a clock. If you require a multi-function unit like the USB-1608GX-2AO that supports hardware paced digital I/O then our USB-1616HS-2 or -4 model will work. If you require just digital support then our USB-DIO32HS will do the job.

16 months ago
Measurement Computing

Was this answer helpful? Yes (1) / No (0)

  • Job Title:
  • Industry:
  • Application:
37 months ago
flaw Houston, TX, USA


Answers

Thank you for your inquiry. The Universal Library will support concurrent paced operation of the analog inputs and the analog outputs on the USB-1608GX-2AO device.

37 months ago
MeasurementComputing

Was this answer helpful? Yes (1) / No (0)

Included Software


  • TracerDAQ® for acquiring and displaying data and generating signals
  • InstaCalTM software utility for installing, configuration, and testing
  • Universal Library includes support for Visual Studio® and Visual Studio®.NET, including examples for Visual C++®, Visual C#®, and Visual Basic®.NET
  • ULx for NI LabVIEWTM Library of VIs and example programs to speed your NI LabVIEW development
  • Universal Library for Android® includes support and examples for the Android 3.1 platform (API level 12) and later
  • Linux® Support for MCC DAQ products

Optional Software


  • DAQamiTM - data acquisition companion software for acquiring data and generating signals
  • DASYLab®2016 Software lets you interactively develop PC-based data acquisition applications by simply attaching functional icons
  • TracerDAQ®Pro - Out-of-the-box virtual instrument suite with strip chart, oscilloscope, function generator, and rate generator – professional version
  • MATLAB® Support for MCC Products - Software Environment for Data Acquisition, Data Analysis, and Application Development

Product Manuals


USB-1608G
document
 
USB-1608G-OEM
document
 
USB-1608GX
document
 
USB-1608GX-2AO
document
 
USB-1608GX-2AO-OEM
document
 
USB-1608GX-OEM
document
 

 

Quick Start Guide


DAQami™ v4.2
QS-MCCDAQ.pdf
 
DASYLab® LITE
QS-DASYLab-MCCDRV.pdf
 
ULx for NI LabVIEW™
QS ULx for NI LabVIEW.pdf
 

 

Product Videos


MCC Video Library
video
 

 

Programming Example Downloads      download

1 Year Warranty

1 Year Warranty

All products are backed with a one year warranty. Buy with confidence.

30 Day Money Back Guarantee

30-Day
Guarantee

If you aren’t happy, simply return the product for a full refund (U.S. customers only).

Live Help

Live Help

Having trouble deciding which product is right for your application? Chat with one of our sales engineers.