I'm trying to make a measurement where I trigger the aquisition. I perform about 12 measurement but with the external trigger the timing jitter should be below 55ns. Therefor it is important to know what the conversion period is.
- Job Title: Hardware designer
- Industry: Medical
- Application: Test tooling
2 days ago
G
Best Netherlands
Answers
Yes, the conversion period is 1 over the sampling frequency. At the maximum sample frequency of 216 kHz, the trigger delay could be up to 4.6 uS. This cannot be reduced or made to be exactly repeatable, since the 24 bit A/Ds are being overclocked by a much faster internal clock that is already running.
3 days ago
Measurement Computing
Was this answer helpful?
Yes (0) /
No
(0)