Data from our DTA instrument (Orton Ceramics) is acquired by PCI DAS 08 card. Card is taken by PC system board. Since recently we are getting following error messages: AIn scan retuened an error=1, Stop Background returned error=1. Does this indicate any problems with DAS08?.
- Job Title: Geologist
- Industry: Government Geological Survey
- Application: Differential Thermal Analysis
75 months ago
Ben56
Colombo, Sri Lanka
Answers
Most likely the PCI-DAS08 is OK. Error 1 was intended to mean that the board is not present in the system, or the logical board number (0, 1, 2, etc.) the program is referencing has no corresponding board configured in our InstaCal utility. Normally you run InstaCal to assign a board number must match what the application programmer expected it to be. If this does not resolve it, then the following FAQ identifies 2 other causes.
http://kb.mccdaq.com/KnowledgebaseArticle50403.aspx
75 months ago
Measurement Computing
Was this answer helpful?
Yes (3) /
No
(0)